ENG

    ATR BR

    ATR BR – Compact Refractometer

    SCHMIDT + HAENSCH ATR BR is a compact, stand-alone critical angle refractometer designed for precise measurement of liquid samples with different viscosities, regardless of their opacity or color.

    Thanks to its robust design, easy operation, and excellent price-performance ratio, the ATR BR is suitable for both laboratory and production applications. Only 0.3 mL of sample is required for stable and precise measurements.

    Category:

    KEY ADVANTAGES

    APPLICATIONS

    TECHNICAL SPECIFICATIONS

    Key Advantages

    • Precise and stable measurements
    • Suitable for low- and high-viscosity samples
    • Measurements independent of sample color and opacity
    • Small sample volume – only 0.3 mL
    • Easy and user-friendly operation
    • Compact and easy-to-clean sample chamber
    • Up to 10 user-defined measuring scales
    • Factory-programmed Brix scale
    • PC connectivity and remote control
    • Form creation and printing
    • Ethernet data transfer
    • Excellent price-performance ratio

    Applications

    The ATR BR is suitable for a wide range of applications in the chemical, food, pharmaceutical, medical, and petrochemical industries.

    It can be used for the analysis of yoghurt, jam, fruit extracts, syrups, honey, coffee extracts, vegetable fats, chocolate, milk, and baby food, as well as for quality control of soft drinks, fruit juices, wine, and beer.

    In the chemical industry, it is suitable for analyzing solvents, organic polymers, distilled products, and inorganic solutions. In petrochemical applications, it can be used for monitoring oils, fats, waxes, crude oil, varnishes, and paints.

    Technical Specifications

    Specification
    ATR BR
    Measuring range
    1,32000–1,54000 RI / 0–100% Brix
    Accuracy
    ±0,00002 RI / ±0,01% Brix*
    Repeatability
    ±0,00001 RI / ±0,01% Brix
    Resolution
    0,00001 RI / 0,01% Brix
    Temperature range
    5–50 °C
    Sample volume
    0,3 mL
    Prism
    Sapphire
    Wavelength
    589 nm
    Display
    Backlit LCD
    Communication
    RS232, parallel interface
    Conformity
    International Pharmacopoeia, ASTM, AOAC, DIN, FDA, ICUMSA

    * Accuracy depends on the measuring range: ±0.00002 RI for 1.32–1.44 and ±0.00003 RI for 1.44–1.54.

    Privacy Overview

    This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.